Intramedullary rodding of femur fractures, although a safe and rapidly performed procedure, can result in several complications. If the rod fit is too loose, fracture instability, rod migration, and delayed union may result. If the rod fit is too tight, cracking of the femur may occur during rod insertion. These complications were investigated in terms of geometric and mechanical parameters of the bone-implant system. Results showed that rods of the same nominal size from different manufacturers showed more than twofold difference in flexural rigidity and a threefold difference in torsional modulus. These differences appear to be due to differences in cross sectional shape and wall thickness of the rods. Measurements of pushout force and hoop stress in cadaver femora showed a large difference in pushout force with different rods, and significantly lower forces in distal than in proximal femoral fracture components. Pushout force decreased with fracture component length proximally and dropped to zero in distal components less than 170 mm long. An increase in ream diameter in the distal components of just 1 mm was found to decrease the mean pushout force from 740N to 90N. The most significant variable was found to be anterior offset of the starting hole more than 6 mm from the centerline of the medullary canal which resulted in consistent lifting of the anterior cortex during insertion of the rod.
Skip Nav Destination
Article navigation
Research Papers
Biomechanical Factors Affecting Fracture Stability and Femoral Bursting in Closed Intramedullary Rod Fixation of Femur Fractures
A. F. Tencer,
A. F. Tencer
Division of Orthopedics, The University of Texas Medical Branch, Galveston, Tex. 77550
Search for other works by this author on:
M. C. Sherman,
M. C. Sherman
Biomedical Engineering Program, The University of Texas at Arlington, Arlington, Tex. 76019
Search for other works by this author on:
K. D. Johnson
K. D. Johnson
Division of Orthopaedic Surgery, The University of Texas Health Science Center at Dallas, Dallas, Tex. 75235
Search for other works by this author on:
A. F. Tencer
Division of Orthopedics, The University of Texas Medical Branch, Galveston, Tex. 77550
M. C. Sherman
Biomedical Engineering Program, The University of Texas at Arlington, Arlington, Tex. 76019
K. D. Johnson
Division of Orthopaedic Surgery, The University of Texas Health Science Center at Dallas, Dallas, Tex. 75235
J Biomech Eng. May 1985, 107(2): 104-111 (8 pages)
Published Online: May 1, 1985
Article history
Received:
March 5, 1984
Revised:
August 1, 1984
Online:
June 15, 2009
Citation
Tencer, A. F., Sherman, M. C., and Johnson, K. D. (May 1, 1985). "Biomechanical Factors Affecting Fracture Stability and Femoral Bursting in Closed Intramedullary Rod Fixation of Femur Fractures." ASME. J Biomech Eng. May 1985; 107(2): 104–111. https://doi.org/10.1115/1.3138530
Download citation file:
Get Email Alerts
Related Articles
A New External Fixator Design for Femoral Fracture Reduction
J. Med. Devices (June,2008)
The Biomechanical Effect of Torsion on Humeral Shaft Repair Techniques for Completed Pathological Fractures
J Biomech Eng (February,2012)
Biomechanical Evaluation of a Novel Lumbosacral Axial Fixation Device
J Biomech Eng (November,2005)
Related Proceedings Papers
Related Chapters
Introduction and Definitions
Handbook on Stiffness & Damping in Mechanical Design
Use of Large Standoff Magnetometry for Geohazard Pipeline Integrity Investigations
Pipeline Integrity Management Under Geohazard Conditions (PIMG)
Mechanics of Distal Locking in Intramedullary Rods: Comparison in Osteoporotic and Normal Femurs
Intramedullary Rods: Clinical Performance and Related Laboratory Testing